CALIBRATION / EVIDENCE WORKSHEET

6-Axis Force Sensor Calibration Certificate Review

Check six-axis uncertainty, find missing evidence and export a review memo. Use a synthetic example or start blank.

Check certificate evidence

Compare all six axes with your limits. This local worksheet uploads no files.

EXAMPLE ONLY — synthetic data, including any edits. Start blank for your own record.
Choose a teaching example
1. Record identity
2. Per-axis expanded uncertainty

Positive finite values only. Full scale is each axis reference magnitude (N or N·m), not the peak-to-peak span. U must be expanded uncertainty at k=2 in % FS; do not enter % of reading or repeatability.

3. Your limits and crosstalk

These are application inputs, not ISO/GUM limits. Equality is included. Compare maximum reported crosstalk using the same receiving-axis FS normalization and relevant applied loads. No error-plus-uncertainty conformity decision is made.

4. Evidence marked by you

“Reported” records your assessment, not independent verification. Unknown and No remain visible gaps. Supplementary records may supply evidence absent from the certificate.

This worksheet does not authenticate documents, verify an accreditation directory, calculate a full uncertainty budget or approve a robot. The report below explains those boundaries.

No current result. Review the example, or start blank and enter your certificate evidence. Editing any field clears the previous result.

1. Four decisions the certificate must support

Compare compatible uncertainties

Use the stated units, coverage factor and reference basis. The worksheet limits are user inputs; GUM does not set a universal robot acceptance percentage.

Source basis: BIPM JCGM 100:2008 — uncertainty evaluation, sections 4–6. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

Separate calibration from acceptance

A release decision needs a specification and an agreed decision rule. Being within the entered worksheet limits is only a numerical comparison.

Source basis: ILAC G8:09/2019 — decision rules and statements of conformity. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

2. Evidence to request before acceptance

Use this engineering procurement checklist alongside the actual report and referenced configuration records. These are review questions, not a claim that every item must appear on every certificate.

Traceability review path: Reference → Calibration → Installed result. Stated uncertainty; Serial + configuration; Fitness still reviewed.ReferenceStated uncertaintyCalibrationSerial + configurationInstalled resultFitness still reviewed
Traceability review path. Conceptual workflow; numerical values appear only in the labeled synthetic example.
2. Evidence to request before acceptance — review worksheet
EvidenceWhat to compareIf missing
Record identityReport ID, sensor model/serial, issuer and calibration dateAsk the issuer for a serial-matched original record.
Measurement chainIntegrated or external electronics, gain, cables and revisionIdentify the configuration actually calibrated.
TraceabilityReference chain and associated uncertaintiesRequest supporting evidence; lack of accreditation is not proof of fraud.
Accreditation claimAccreditation ID and relevant quantity/range/method in the body directoryKeep scope unverified; a generic logo is insufficient.
Calibration configurationMatrix or internal calibration revision; raw vs engineering-unit outputClarify processing location and serial-specific version.
Results and conditionsPer-axis results, units, load directions, uncertainty basis and relevant mounting/environmentRequest omitted data or document restricted use.

Source basis: ISO/IEC 17025:2017 — laboratory competence (official scope) · ATI — Force/Torque Sensor FAQ, 2020, sections 4.3–4.4 (matched systems and matrix processing) · NIST — policy on metrological traceability and fitness for purpose. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

3. Matrix revision, units and reference frame

For a six-input linear model, w = C(s − s₀). Coefficients may be applied in the interface or host. Keep the calibration ID, channel order, zero handling and engineering units consistent; avoid applying the matrix twice.

Process the calibration once: Raw channels → Matched calibration → Wrench + frame. Zero + input units; Interface OR host; N and N·m.Raw channelsZero + input unitsMatched calibrationInterface OR hostWrench + frameN and N·m
Process the calibration once. Conceptual workflow; numerical values appear only in the labeled synthetic example.
3. Matrix revision, units and reference frame — review worksheet
Review itemRequired interpretationBoundary
RevisionMatch the physical unit and active calibration, not just the model nameDigital products may expose an internal ID rather than a public matrix.
Origin and signsState the frame and use M = r × F for an origin shiftAn origin change alters moments even when the sensor is healthy.
Condition numberCompare only after defining input/output scaling and unitsNo universal κ < 15 or κ > 50 quality cutoff; N vs kN changes an unnormalized matrix.
Higher-order modelSelect on held-out residuals across the intended load domainComplexity adds calibration effort and extrapolation risk; no universal percentage improvement.

Source basis: ATI — Force/Torque Sensor FAQ, 2020, sections 4.3–4.4 (matched systems and matrix processing). Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

4. Worked uncertainty comparison: what the tool computes

Synthetic example: Fz full scale = 600 N and expanded U = 0.42% FS give U = 2.52 N. At a 20 N working load this equals 12.6% of reading. A small % FS can therefore be unsuitable at low load. This is a unit conversion, not a new uncertainty budget.

Synthetic Fz uncertainty conversion: 600 N full scale → × 0.42 / 100 → 2.52 N expanded U. Not peak-to-peak; Entered % FS; 12.6% at 20 N.600 N full scaleNot peak-to-peak× 0.42 / 100Entered % FS2.52 N expanded U12.6% at 20 N
Synthetic Fz uncertainty conversion. Conceptual workflow; numerical values appear only in the labeled synthetic example.
4. Worked uncertainty comparison: what the tool computes — review worksheet
Synthetic input or resultValueMeaning
Fz reference magnitude600 NPositive FS magnitude; not the ±600 N peak-to-peak span.
Reported expanded U0.42% FS, k=2Converted absolute U = 600 × 0.42 / 100 = 2.52 N.
Entered force U limit0.50% FS0.42 ≤ 0.50: within this entered U limit.
Working load comparison2.52 / 20 × 100 = 12.6%Relative uncertainty at 20 N; application suitability remains a separate decision.
Coveragek=2; approximately 95% only under suitable assumptionsSmall degrees of freedom or non-normal distributions need another treatment.

Source basis: BIPM JCGM 100:2008 — uncertainty evaluation, sections 4–6. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

5. Uncertainty budget: evidence beyond a single number

Type A and Type B describe evaluation methods, not a random-versus-systematic split. A budget needs standard uncertainties, sensitivity coefficients and correlations. Do not add datasheet maxima as if they were independent standard deviations.

Budget review, not a worksheet calculation: Standard inputs → Model + covariance → U = k × u_c. Type A / Type B; Avoid double counting; State coverage basis.Standard inputsType A / Type BModel + covarianceAvoid double countingU = k × u_cState coverage basis
Budget review, not a worksheet calculation. Conceptual workflow; numerical values appear only in the labeled synthetic example.
5. Uncertainty budget: evidence beyond a single number — review worksheet
ContributionWhat to requestKnown / unknown in this worksheet
Reference force and lever armReference uncertainty, geometry and sensitivity modelUnknown: the worksheet does not evaluate reference equipment.
Repeatability and reversibilityRepeated readings and direction-dependent residualsUnknown: a single stated U does not expose repeat data.
Temperature, mounting and electronicsInfluence coefficients, actual conditions and correlation treatmentUnknown: installed influences may differ from calibration.
CombinationFor independent contributions, u_c² = Σ(cᵢuᵢ)²; otherwise include covarianceNot calculated here; source budget must justify its model.
ExpansionU = k × u_c, with stated k and coverage interpretationEntered U is compared only after the user confirms its basis.

Source basis: BIPM JCGM 100:2008 — uncertainty evaluation, sections 4–6. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

6. What standards and accreditation do establish

Read the actual scope and contracted method. A uniaxial force standard does not automatically supply a six-axis combined-load protocol. No universal number of points, U limit, bolt torque or annual interval is asserted here.

6. What standards and accreditation do establish — review worksheet
ReferenceRelevant scopeDo not infer
ISO/IEC 17025:2017Competence, impartiality and consistent laboratory operationAccreditation of this site or fitness of every sensor application.
ISO 376:2011Calibration/classification for static uniaxial testing-machine force-proving instrumentsValidation of a full six-component combined-load envelope.
JCGM 100:2008Framework for evaluating and expressing measurement uncertaintyA mandatory 0.5% FS robot acceptance limit.
ILAC G8:09/2019Decision rules and statements of conformityThat comparing U alone proves error-plus-uncertainty conformity.

Source basis: ISO/IEC 17025:2017 — laboratory competence (official scope) · ISO 376:2011 — official uniaxial force-proving scope · ILAC G8:09/2019 — decision rules and statements of conformity. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

Ready to request missing evidence? Download the blank certificate checklist

7. Mounting, measurement scope and cost tradeoffs

Use the manufacturer drawing and laboratory procedure for fasteners, surfaces and environmental conditions. The numerical installation sensitivity of a specific sensor is unknown without model-specific evidence.

Compare the measurement configuration: Lab setup → Installation changes → Verification plan. Mounting + electronics; Record differences; No universal correction.Lab setupMounting + electronicsInstallation changesRecord differencesVerification planNo universal correction
Compare the measurement configuration. Conceptual workflow; numerical values appear only in the labeled synthetic example.
7. Mounting, measurement scope and cost tradeoffs — review worksheet
Risk / tradeoffDecision impactSmallest useful next step
Misuse: same-family spare with old coefficientsA functioning output is not proof of the correct calibrationMatch the active serial/calibration ID and recheck the measurement chain.
Scenario mismatch: single-axis evidence for combined useUnverified residuals at the actual force/moment combinationAsk for relevant combined-load results or restrict the operating domain.
Installation mismatch: changed flange or amplifierCalibration conditions no longer describe the installed systemLog the change and agree a verification plan; do not guess a universal correction.
Cost: broader calibration scopeMore load cases, fixture work, transport and downtime may be neededQuote only necessary ranges/directions first; published lead time and cost are unknown.
Alternative: restricted use pending evidenceMay reduce immediate rework but limits valid decisionsHave the application owner document the restriction and unresolved uncertainty.

Source basis: ATI — Force/Torque Sensor FAQ, 2020, sections 4.3–4.4 (matched systems and matrix processing) · NIST — policy on metrological traceability and fitness for purpose. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

8. Digital calibration certificates: format is not authenticity

The cited PTB documentation is version 3.3.0, not a claim about the latest release. It defines XML structure; signatures are optional. Schema validation and signature verification are different checks. This page performs neither.

Three separate digital evidence checks: XML schema → Signature if present → Scope + identity. Structure only; Trust chain required; Engineering review.XML schemaStructure onlySignature if presentTrust chain requiredScope + identityEngineering review
Three separate digital evidence checks. Conceptual workflow; numerical values appear only in the labeled synthetic example.
8. Digital calibration certificates: format is not authenticity — review worksheet
CheckPDF or structured dataRemaining review
Document formatPDF may be an original; PTB DCC uses versioned XMLJSON-LD page metadata is not a PTB DCC.
Structural validityAn XML file can be checked against its stated XSDValid structure does not prove truthful measurement results.
SignatureVerify signature and trust chain when presentDCC does not inherently require a signature or guarantee accreditation.
Software ingestionCheck axis order, units, serial and revision before importAutomation reduces transcription work but still needs mapping and review.

Source basis: PTB DCC schema 3.3.0 — versioned XML documentation and optional signatures. Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

9. Gravity checks: useful, but not independent six-axis calibration

Engineering derivation: at rest, F = m·g in sensor coordinates and M = r × F. With a fixed center-of-mass vector r, [F; M] = [I; [r]×]F spans at most three dimensions. More poses do not create six independent load inputs. For r = (0, 0, L), Mz is zero in every static pose.

Fixed center of mass limits gravity verification: F = m·g_s → M = r × F → At most rank 3. Static sensor frame; Fixed r couples loads; Axial r gives Mz = 0.F = m·g_sStatic sensor frameM = r × FFixed r couples loadsAt most rank 3Axial r gives Mz = 0
Fixed center of mass limits gravity verification. Conceptual workflow; numerical values appear only in the labeled synthetic example.
9. Gravity checks: useful, but not independent six-axis calibration — review worksheet
Illustrative static sensor-frame forceMoment for r = (0, 0, L)What it can check
F = (0, 0, +mg)M = (0, 0, 0)An axial response at this load; sign depends on the declared convention.
F = (+mg, 0, 0)M = (0, +Lmg, 0)Coupled Fx/My response, not independent pure Fx and My gains.
F = (0, +mg, 0)M = (−Lmg, 0, 0)Coupled Fy/Mx response; no axial Mz excitation.
Additional fixed-r posesAlways M = r × FUseful consistency checks; cannot independently identify all matrix terms.
Independent Mz verificationRequires a known nonzero axial torqueUse an appropriate reference torque or additional independent load fixtures.

Basis: static wrench equations shown above. This derivation describes observability, not a validated field calibration protocol.

10. Recalibration and in-service review

Intervals are a quality-system decision based on history, use, manufacturer guidance and risk. A calibration date is not a universal expiry date. Any suspected damaging event warrants review without waiting for an arbitrary overload percentage.

Evidence-based review after a change: Change / event → Reference checks → Document decision. Log configuration; Investigate mismatch; Review scope + interval.Change / eventLog configurationReference checksInvestigate mismatchDocument decisionReview scope + interval
Evidence-based review after a change. Conceptual workflow; numerical values appear only in the labeled synthetic example.
10. Recalibration and in-service review — review worksheet
TriggerImmediate reviewNext decision
Collision, overload or unexplained driftRecord event and isolate questionable measurements under site procedureInspect and agree reference checks, recalibration or replacement with the responsible engineer.
Electronics, firmware, matrix or mechanical changeCompare configuration against the calibrated chainDetermine what verification or recalibration is needed for that design.
Routine interval reviewUse drift history, usage and criticalityDocument an evidence-based interval; no universal 12-month mandate.

Source basis: ILAC G24:2022 — determination and review of recalibration intervals (G-series directory). Engineering recommendations are the site’s interpretation; no source endorses this worksheet.

11. Three worked decision scenarios

Illustrative teaching scenarios, not customer case histories or measured product results.

Example A: complete numbers, unknown accreditation scope

Premise: The preloaded synthetic record has all six expanded U values and user-entered limits.

Process: Run the worksheet: Fz is 0.42% FS, Mz is 0.52% FS, and the scope question remains Unknown.

Outcome / next step: Numbers are within entered limits; the scope gap remains. Request the relevant accreditation scope if the procurement task requires it.

Example B: missing calibration revision

Premise: Load the missing-evidence example; its matrix revision is blank.

Process: Run the same numerical comparison. Matrix, traceability and combined-load evidence remain gaps even if all numbers are small.

Outcome / next step: Request the serial-matched calibration configuration. Filling a revision only resolves that gap when the matching evidence is also marked checked.

Example C: same % FS, different working-load decision

Premise: A synthetic Fz axis has 600 N FS and U = 0.42% FS; the application works at 20 N.

Process: Convert to 2.52 N expanded U, or 12.6% of the working reading.

Outcome / next step: A 0.5% FS worksheet limit can be met while the process requirement is not. Review working-load uncertainty and the application decision rule.

Reproduce the worksheet result

Choose Load worked example, then Review evidence. The result retains one unknown scope item even when the numbers are within the entered limits.

6-axis force sensor calibration certificate worksheet showing synthetic per-axis uncertainty comparisons and one unresolved evidence item
Actual worksheet execution captured November 4, 2026. Synthetic inputs: FS Fx/Fy/Fz = 300/300/600 N; Mx/My/Mz = 15/15/15 N·m. Expanded U = 0.35/0.38/0.42/0.48/0.50/0.52% FS (k=2); force/moment limits = 0.50/0.60% FS; crosstalk = 1.15%, limit = 1.50%. Scope is Unknown; the other ten evidence items are marked reported. The result card is isolated from site navigation. This is software evidence, not a physical calibration, customer record or product approval.

12. Questions by review stage

Evidence

What should a 6-axis force sensor calibration certificate tell me?

Which unit and configuration were measured, the results, units, relevant conditions and associated uncertainty. Matrix/frame details may be in referenced configuration records. Use the checklist to request missing evidence.

Does this worksheet generate a calibration certificate?

No. It exports a user-entered evidence review memo. No physical test, issuer authentication or accreditation validation is performed.

Can I use the preloaded example as my certificate?

No. All example identifiers and values are synthetic. Edits retain the example label. Choose Start blank review for your own record.

Does an accreditation logo prove the report covers my test?

No. Check the accreditation body directory and the relevant quantity, range and method. This website does not verify that directory.

Is an unaccredited factory report necessarily invalid?

No. Examine its method, traceability, uncertainty and suitability for the contracted task. Accreditation may be required by your procurement or quality procedure.

Must the cable and interface always have the same serial?

This is design-specific. ATI documents matched systems, while other designs may integrate conversion inside the sensor. Record the calibrated configuration and follow its manufacturer instructions.

Uncertainty

Why enter six uncertainty values?

Axis ranges and uncertainties can differ. A single force number can hide the worst axis; the result shows each entered axis and converts its uncertainty to N or N·m.

Is k=2 always exactly 95% coverage?

No. That approximation depends on the distribution and effective degrees of freedom. Use the coverage interpretation stated by the laboratory.

Can I enter percent of reading as percent FS?

No. Convert against the explicit reference magnitude at the relevant load first. If the basis is unclear, request clarification rather than inventing a number.

Does Within entered limits mean accepted?

No. It compares reported U and crosstalk with your inputs, including equality. It does not evaluate measured error against a tolerance with an agreed conformity decision rule.

Do filtering and tare remove calibration uncertainty?

No. They can affect noise and zero but do not automatically remove gain, reference, hysteresis or model uncertainty. Reassess the measurement model and conditions.

How is crosstalk normalized here?

Use the absolute maximum residual in percent of the affected receiving axis full scale at the relevant applied loads. Other definitions require conversion and an agreed test method.

Integration

Must a certificate publish all 36 matrix elements?

Not necessarily. An integrated sensor may use protected internal coefficients. You still need a traceable calibration identifier, configuration, units, frame and relevant results.

Can a fixed-payload gravity test validate Mz independently?

Not when the center of mass is on the sensor z-axis: Mz = rxFy − ryFx = 0. With any fixed center of mass, moments remain coupled to force. Use independent known loads for independent channel validation.

Does an electrical shunt check replace mechanical calibration?

No. Simulating a bridge signal does not apply known mechanical force or torque and cannot establish the full mechanical response.

Is there one correct mounting torque for all sensors?

No. Follow the specific drawing and procedure, including bolt and thread conditions. This guide supplies no universal fastener setting.

Lifecycle

Does a DCC file guarantee authenticity?

No. PTB defines structured XML; a signature is optional in the cited version. Validate the format, any signature and its trust chain, and the report scope separately.

How often should I recalibrate?

Set and review an interval from drift history, usage, manufacturer advice and your quality procedure. Investigate suspected damage or configuration changes separately; no fixed interval fits every task.

13. Sources, review date and limitations

Prepared and reviewed by HexaForce Metrology & Robotics. Repository record created ; public publication date is unverified. Review again by May 4, 2027, or sooner if the cited methods change.

Source and content review: . Editions are identified where cited; the PTB schema link deliberately targets version 3.3.0. Official scope pages do not replace the full standards or the laboratory’s contracted method.

No model-specific calibration records, laboratory accreditation evidence, service lead times or prices were supplied for this guide. Those remain unknown. Examples and limits are illustrative; worksheet entries are unverified. The worksheet performs input checks, evidence counting, % FS conversion and comparisons only.

Download the blank evidence checklist (CSV) or return to the local review tool. For the wider calibration method, see 6-axis load cell design and calibration.

Resolve the gaps before accepting the report

Send the certificate ID, sensor serial, measurement-chain details and unresolved evidence list. Use the exported memo for your requirements; the worksheet does not authenticate certificates.

Continue with axis and sign checks, acquisition-chain review, or engineering request resources.

Request certificate evidence review