6-Axis Force Sensor Calibration Laboratory Selection

Screen a candidate lab and generate an RFP checklist. Compare scope, uncertainty, loading and readout claims, then verify the evidence before contracting.

Calibration Lab Screening & RFP Worksheet

Illustrative defaults, not verified laboratory claims. All numeric fields are required; the reference ID is optional. Replace the example inputs before screening a real candidate. Unknown claims need written evidence. Scoring method and limits.

Choose an illustrative candidate laboratory profile
1. Sensor Specification & Tolerance Demands
0.01–5% FS; one common limit for all axes. If tolerances differ, use the strictest limit for a conservative screen or evaluate each axis separately.
2. Pillar 1: Rig Architecture & Loading Vectors
Whole number: 1–500; count alone does not prove load-case coverage
3. Pillar 2: Scope & Achievable Calibration Uncertainty
Use the worst per-axis U for Fx/Fy/Fz; 0.001–5% FS, approximately 95% coverage
Use the worst per-axis U for Mx/My/Mz; 0.001–5% FS, approximately 95% coverage
Leave unchecked if unknown: TUR is withheld and the RFP asks for clarification. Convert % of reading or absolute units first; CMC may understate the actual sensor calibration uncertainty. See worked conversions.
4. Pillar 3: Traceability Chain & Primary Standards
5. Pillar 4: Decoupling Matrix & Crosstalk
6. Pillar 5: Logistics, Fixturing & Turnaround
Awaiting evaluation. Use the illustrative defaults or enter candidate claims, then click Evaluate. Unknown uncertainty basis still produces an RFP checklist.

Six Decisions Before Requesting a Quote

Verify the contracted work, not the logo

Match quantities, load directions, ranges and method to the schedule. Ask which results will carry accreditation and which will be outside scope.

Source basis: ILAC G18:01/2024 — describing scopes of accreditation. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

Compare like uncertainty quantities

Use per-axis expanded uncertainty at the intended points, not an unconverted % of reading or the smallest CMC on a brochure.

Source basis: ILAC P14:09/2020 — uncertainty in calibration. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

Agree the output before testing

A measured value with uncertainty differs from a conformity statement. Specify the decision rule if pass/fail is required.

Source basis: ILAC G8:09/2019 — decision rules and statements of conformity. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

Keep the reference chain auditable

Obtain suitable reference certificates and uncertainties. An accredited transfer-standard route can be appropriate without a direct NMI certificate.

Source basis: ILAC P10:07/2020 — metrological traceability. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

Validate the complete measurement chain

Agree which matrix, electronics, firmware and fixtures are included. A new 6x6 matrix is a possible deliverable, not mandatory for every recalibration.

Source basis: ISO/IEC 17025:2017 — laboratory competence. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

1. Loading methods: select coverage, not a machine label

Original procurement comparison: no architecture is automatically qualified. Record the applied wrench and all output channels, then check repeatability and independent application-relevant cases.

Select load coverage
  1. Define load cases
  2. Record six outputs
  3. Validate residuals

Sequential and combined methods need evidence appropriate to their intended use.

1. Loading methods: select coverage, not a machine label — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Combined-load frameAchievable vectors, alignment and reference uncertaintyValidate interactions at selected combined statesA hexapod name does not establish accuracy
Sequential loadingIndependent signed load cases; six outputs per case; remounting checksIdentify linear sensitivities where the load basis has sufficient rankDoes not by itself validate all combined-load nonlinearities
Deadweight/lever fixturesForce, geometry and fixture uncertainty; directions coveredMay realize individual or combined components with suitable fixturesA bare 1D stack alone does not cover six components

Source basis: Schwind (2010) — Multicomponent-transducers: definition, construction types and calibration considerations · Ogushi, Baumgarten & Röske (2015) — crosstalk matrix of a force-torque transducer. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

2. Verify accreditation coverage

Read the current schedule and the quotation together. A uniaxial scope is insufficient evidence of accredited coverage for the complete requested service; it does not prove the laboratory is technically incapable.

Map the contracted service
  1. Find current schedule
  2. Match method and range
  3. Mark coverage gaps

Accreditation status applies to covered operations, not every service the provider offers.

2. Verify accreditation coverage — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Identity and statusAccreditation body directory entry, site address, schedule revisionConfirm the actual facility and current statusMarketing logos are insufficient
Work coverageQuantities, ranges, directions, methods and exclusionsMark each quoted operation as covered, outside scope or unknownDo not require one specific phrase such as “6-DOF”
Report and subcontractingSample report; scope status of subcontracted workAgree which outputs are issued under accreditationA single-axis accredited report is not a full-system endorsement

Source basis: ILAC G18:01/2024 — describing scopes of accreditation · ISO/IEC 17025:2017 — laboratory competence. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

3. Traceability: NMI and accredited transfer routes

The worksheet requests documentary evidence rather than ranking laboratories by distance from an NMI. Suitability depends on the reference service, range and uncertainty.

Follow the reference evidence
  1. Reference certificates
  2. Suitable uncertainty
  3. Working load model

NMI and suitable accredited services can provide traceability routes.

3. Traceability: NMI and accredited transfer routes — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
NMI serviceCertificate, covered service and uncertaintyCheck suitability for the required rangeAn NMI name alone is not a complete chain
Accredited transfer calibrationReference certificates and relevant scopeA valid route where appropriate for intended useReview drift and intermediate checks
Working measurement chainReference IDs, corrections and uncertainty contributionsConnect the applied loads to reported componentsNo universal percentage is assumed

Source basis: ILAC P10:07/2020 — metrological traceability · EURAMET cg-4, version 3.0 (02/2022) — uncertainty of force measurements. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

4. Matrix and residual validation

Decide whether the job verifies the existing calibration or generates new coefficients. The following checks are original engineering recommendations; no universal condition-number threshold is asserted.

Validate the intended model
  1. Identify revision
  2. Fit or verify model
  3. Independent checks

Condition numbers require stated units, normalization and norm; no universal cutoff is used.

4. Matrix and residual validation — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Existing matrix verificationRevision, units, channel order and independent residualsRetain the matrix if evidence meets agreed requirementsA coefficient file may remain proprietary
New linear/nonlinear modelFit data, independent validation and model rangeChoose model complexity using validation performanceMore terms alone do not prove better calibration
Condition numberNormalization, units, norm and singular valuesInvestigate sensitivity in the stated coordinate scalingChanging force/moment units changes an unnormalized value

Source basis: Schwind (2010) — Multicomponent-transducers: definition, construction types and calibration considerations · Ogushi, Baumgarten & Röske (2015) — crosstalk matrix of a force-torque transducer. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

5. TUR: a comparison, not a verdict

For symmetric limits ±T, this worksheet uses TUR = T/U, equivalent to tolerance span/(2U). Each example uses quoted achievable expanded uncertainty U on the same per-axis full-scale basis at approximately 95% coverage. No measured error is entered, so the tool cannot decide conformity or false-accept probability.

Compare uncertainties before deciding
  1. Same per-axis basis
  2. Calculate T / U
  3. Agree decision rule

TUR = T/U for symmetric ±T limits. No measured error means no conformity verdict.

5. TUR: a comparison, not a verdict — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Illustrative force exampleFS = 1000 N; T = 0.20% FS = 2 N; U = 0.05% FS = 0.5 NTUR = 4:1; compare with the agreed ruleThis is not evidence that the sensor passes
Illustrative torque exampleFS = 50 N·m; T = 0.20% FS = 0.10 N·m; U = 0.05% FS = 0.025 N·mTUR = 4:1 on this axis onlyEach other axis and load point needs review
Reading-based uncertaintyU = 0.05% of 100 N = 0.05 N; FS = 1000 NConvert to 0.005% FS before comparisonInclude all actual uncertainty terms first
Decision ruleAgree acceptance limits and risk treatment in writingA guard band such as g = U is one possible rule4:1 is an editorial target here, not a universal ISO rule

Source basis: ILAC G8:09/2019 — decision rules and statements of conformity · ILAC P14:09/2020 — uncertainty in calibration · JCGM 106:2012 — measurement uncertainty in conformity assessment. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

6. Combined loads and safe test coverage

Use the sensor maker’s approved combined-load envelope and the laboratory’s fixture limits. This page does not assume an ellipsoid, independent full-scale simultaneous loads or safe overload levels.

Define a safe load plan
  1. Approved limits
  2. Application vectors
  3. Validate selected states

Use approved sensor and fixture envelopes; no generic ellipsoid is assumed.

6. Combined loads and safe test coverage — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Application statesRepresentative six-component wrench vectors and duty conditionsSelect independent validation casesStatic evidence does not establish dynamic bandwidth
Sensor/fixture limitsApproved combined-load and mounting limitsAgree a safe test plan before loadingSingle-axis ratings do not define combined safety
Coverage gapsUntested directions, low loads, temperatures and combinationsMark these as unknown in the reportDo not extrapolate a fit beyond evidence

Source basis: Schwind (2010) — Multicomponent-transducers: definition, construction types and calibration considerations. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

Turn missing evidence into laboratory questions

Use the worksheet to prepare a tailored RFP, then request the current scope and actual per-axis uncertainty from the candidate laboratory.

Generate an RFP checklist ↗

7. Moment generation and geometry

For a force applied at offset r, M = r × F. An effective lever arm and parasitic components need review when that generation method is used; a calibrated torque reference may use another realization.

Review moment realization
  1. Applied force F
  2. Effective offset r
  3. Moment M = r × F

Include alignment, geometry and deflection or the appropriate torque-reference uncertainty.

7. Moment generation and geometry — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Effective lengthLength realization, alignment and loaded geometryRelate force to the intended moment componentA CMM is one possible method, not a requirement
Cross componentsFull applied wrench and fixture reactionsSeparate intended moments from accompanying forcesA single off-axis force is not a pure couple
UncertaintyReference force, geometry, deformation and correlationsEvaluate actual moment uncertaintyNo fixed arm-error percentage is assumed

Source basis: EURAMET cg-14, version 2.0 (03/2011) — static torque calibration · Schwind (2010) — Multicomponent-transducers: definition, construction types and calibration considerations. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

8. Include the actual readout chain

Procurement checklist: ask for a demonstrated acquisition path for the specified device revision. The examples do not claim tested HexaForce protocol support.

Agree the readout chain
  1. Device and firmware
  2. Acquisition settings
  3. Reported components

Confirm actual device support; a protocol name alone is insufficient.

8. Include the actual readout chain — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Digital outputProtocol variant, device files, firmware, units and matrixConfirm readings can be acquired without changing the calibrated chainEtherCAT, CANopen and RS-485 are not interchangeable
Bridge/analog outputExcitation, conditioning, gain, resolution and cablingInclude the agreed readout configuration in calibrationAnalog voltage capability does not alone prove bridge support
Timing and processingAveraging, filters, sample timing and reference synchronizationMatch the method to static or dynamic needsPTP and 1 kHz are not universal requirements

Original illustrative engineering interpretation; see the limitations below.

9. Fixtures and mounting repeatability

Agree an assembly procedure and verify its effect. This guide does not prescribe bolt grade, preload, material or micrometre limits without the sensor drawing and method evidence.

Control mounting and environment
  1. Assembly procedure
  2. Repeatability checks
  3. Logged conditions

Use drawing- and method-specific limits rather than arbitrary bolt or soak requirements.

9. Fixtures and mounting repeatability — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Interface definitionDrawing, origin, orientation and approved fastener procedureMake the mounting reproducibleNominal robot-flange compatibility is insufficient
Fixture behaviorAlignment, deflection and cable-force checksInclude material effects in the uncertainty modelSteel alone does not guarantee a suitable fixture
EnvironmentTemperature limits, stabilization criterion and logged conditionsUse a method-specific stabilization procedure24 hours at 20°C is not universally required

Original illustrative engineering interpretation; see the limitations below.

10. Turnaround and return-to-service planning

Original scheduling worksheet: quoted weeks in the tool are planning inputs, not market benchmarks. Request written start/end milestones and a contingency for failed intake or adjustment.

Plan the complete service
  1. Intake and fixtures
  2. As-found / as-left
  3. Return verification

Quoted weeks are planning inputs, not delivery promises.

10. Turnaround and return-to-service planning — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Before dispatchScope agreement, adapters, software and packingAvoid queue time spent resolving missing hardwareLead time remains quote-specific
During serviceAs-found data; authorization before adjustment; as-left dataPreserve drift history and change controlDo not overwrite an existing matrix without consent
After returnConfiguration check and application verificationPlan installation and production release separatelyA laboratory report does not validate the whole robot

Source basis: ISO/IEC 17025:2017 — laboratory competence. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

11. Compare total cost and unresolved risk

Original RFQ comparison template: public evidence is insufficient for universal prices, downtime savings or false-accept rates. Enter quoted amounts rather than adopting invented market ranges.

Compare complete quotations
  1. Service and fixtures
  2. Freight and downtime
  3. Evidence still missing

Price and operational cost remain unknown until quoted and evaluated for the site.

11. Compare total cost and unresolved risk — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
Direct service costCalibration, fixtures, matrix work and report chargesCompare the same deliverablesPrice unknown until quoted
Operational costFreight, customs, spares and internal validation timeBudget a realistic return-to-service windowDowntime value is site-specific
Scope mismatchAdditional provider or separate non-accredited workChoose split service only with explicit responsibilitiesLowest quote may omit required evidence

Source basis: ISO/IEC 17025:2017 — laboratory competence. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

12. Understand and use the worksheet score

The score is an original HexaForce editorial rubric. It organizes questions, not metrological acceptance. All entries remain unverified claims and actual laboratory range is not entered or automatically matched.

Use scores as screening aids
  1. Entered claims
  2. Editorial rubric
  3. Request evidence

Any critical gap takes priority over a favorable score. Qualification remains unverified.

12. Understand and use the worksheet score — technical reference criteria
Review itemEvidence to requestDecision useRemaining limit
WeightsLoad coverage 25%; scope/uncertainty 25%; traceability 20%; model validation 15%; logistics 15%See weighted pillar contributions in the resultWeights have no accreditation authority
Ratio bands≥4, ≥3, ≥1.5 and <1.5; unknown if basis unconfirmedIllustrative screening priorities onlyNo probability, guard band or conformity is computed
Next-action gatesAny critical gap → resolve first; otherwise <80, unknown basis, marginal TUR or unconfirmed references → clarify; remaining ≥80 → evidence reviewCopy the tailored RFP and request source documentsA high score never means the laboratory is qualified

Source basis: ILAC G8:09/2019 — decision rules and statements of conformity · ILAC P14:09/2020 — uncertainty in calibration. Tables and decision flows are original procurement interpretations. Sources support the stated principles, not the editorial weights or any laboratory endorsement.

13. Five Illustrative Procurement Scenarios

Illustrative procurement and quality scenarios evaluated using this screening framework. Inputs and outcomes are hypothetical; these are not customer projects, verified product capabilities or measured savings.

Calibration laboratory selection worksheet: score 88, clarifications required before shortlisting
Original browser capture, November 4, 2026, using the unchanged illustrative defaults: Fz 5000 N, Fx/Fy 2500 N, Mz/Mx/My 200 N·m; tolerance 0.30% FS; claimed force U 0.05% FS and torque U 0.08% FS. The uncertainty basis is unconfirmed, so both TUR values remain “Not comparable” despite the 88/100 editorial score. Other inputs: Cartesian rig, combined loading, 36 points, claimed multi-axis scope, transfer references with an NMI certificate, documented moment realization, linear 6x6 model, crosstalk validation, coefficient export, native digital/analog readout, standard flange adapters, three weeks, expedited service and documented stabilization. This capture shows software execution only; it is not a laboratory test, customer endorsement or evidence of physical capability.

Wrist sensor: comparable uncertainty

Operational Premise: Illustrative: Fz FS 1000 N, Mz FS 50 N·m, tolerance 0.20% FS on each axis.

Audit Process & Technical Discovery: Quote uses U = 0.05% FS at approximately 95% coverage; both example ratios are 4:1. Scope and achievable uncertainty at all other axes remain unverified.

Procurement Outcome & Protocol: Shortlist for document review only. Request actual per-point values, fixture/readout scope and the decision rule before contracting.

Ankle sensor: uncertain percentage basis

Operational Premise: Illustrative: an ankle sensor quote states “0.05% uncertainty” without denominator or coverage.

Audit Process & Technical Discovery: Leave the basis confirmation unchecked. The worksheet suppresses TUR and generates uncertainty questions while preserving the other screening results.

Procurement Outcome & Protocol: Ask for units, coverage and per-axis values. Do not convert the claim into a conformity statement.

Sequential calibration: examine the method

Operational Premise: Illustrative: a candidate loads axes sequentially and records all six outputs.

Audit Process & Technical Discovery: Ask for independent signed load cases, load-matrix rank, remounting repeatability and combined-load validation where relevant.

Procurement Outcome & Protocol: Do not reject the laboratory from its rig name. Require evidence for the intended range and residuals.

Digital readout: resolve an acquisition gap

Operational Premise: Illustrative: an EtherCAT device is offered to a laboratory claiming analog-only acquisition.

Audit Process & Technical Discovery: The tool flags Electrical Mismatch even if other claims score well. Ask about a compatible master or a customer-supplied readout.

Procurement Outcome & Protocol: Hold scheduling until the acquisition path, firmware and calibration configuration are agreed.

Semiconductor process: low-load requirements

Operational Premise: Illustrative: process forces are well below sensor full scale; no specific commercial device is claimed.

Audit Process & Technical Discovery: Evaluate actual uncertainty and noise at the operating points, including mounting and readout. A favorable full-scale ratio can hide poor relative performance at low load.

Procurement Outcome & Protocol: Request low-load validation points and environmental conditions; do not infer application fitness from the headline score.

14. Frequently Asked Questions

Accreditation & Scope Verification

Does an ISO/IEC 17025 logo prove six-axis coverage?

No. Check the current schedule and the contracted operations, ranges and methods with the accreditation body.

Must the schedule literally say “6-DOF”?

No. Scope wording varies. Ask the laboratory to map the requested work to its covered quantities and methods.

Is work outside scope necessarily invalid?

No. It is not established as accredited by that scope. Obtain technical evidence and confirm whether the customer permits it.

Can an ISO 9001 certificate replace calibration accreditation?

It does not establish the specific measurement competence and uncertainty of the requested calibration service.

Does the worksheet verify a real laboratory?

No. All inputs are claims. The result is a document-request and comparison aid with no accreditation check or endorsement.

Loading Rigs & Mechanics

Is a hexapod mandatory?

No. Select a method whose reference loads, uncertainty and validation cover the application.

Can sequential loads identify crosstalk?

They can identify a linear cross-sensitivity model with sufficiently independent load cases and all outputs recorded. They do not alone validate nonlinear interactions under every combined state.

Are 36 load points required for 36 coefficients?

No universal count applies. Each case can provide several output observations; rank, repeats, offsets, model complexity and validation matter.

Should all axes be loaded to full scale together?

Only if expressly permitted by the sensor and fixture combined-load limits. Single-axis ratings alone do not establish safety.

Does static calibration establish dynamic accuracy?

No. Dynamic behavior needs an appropriate separate method and evidence.

Traceability & Uncertainty

Is CMC the uncertainty my sensor will receive?

Not necessarily. Request achievable calibration uncertainty including the device and configuration contributions.

What does TUR mean here?

For symmetric ±T limits it is T/U. Both quantities must use the same per-axis full-scale denominator; U is expanded at approximately 95% coverage.

Does 4:1 guarantee a pass?

No. There is no measured error in this tool. Conformity also requires an agreed rule and relevant measurement results.

Is a direct NMI certificate mandatory?

Not universally. Suitable accredited calibration services are also a recognized traceability route.

Is there a universal matrix condition-number limit?

No limit is claimed here. Norm, normalization, units and the model must be defined before interpreting a condition number.

Logistics, Turnaround & Fieldbuses

Must the lab generate a new matrix?

Not for every job. Verification of the existing revision may be the intended service. Agree adjustments and coefficient ownership first.

Must the lab use 1 kHz sampling and PTP?

No universal requirement is assumed. Specify timing, filters and acquisition appropriate to the calibration method.

Is a 24-hour thermal soak compulsory?

Not as a general rule. Agree sensor- and method-specific environmental limits and evidence of stabilization.

What if uncertainty or a capability is unknown?

Do not substitute the preset as evidence. Leave the basis confirmation unchecked, use conservative capability selections and request the missing information through the generated questions.

What is the next step after a favorable score?

Verify the current schedule, achievable uncertainty, sample report, load plan and readout configuration before purchasing the service.

15. Sources, Review Date and Limitations

Authored and reviewed by HexaForce Metrology & Robotics. Technical guide formally established ; sources checked and illustrative model reviewed .

Public source identities and available guidance were checked on November 4, 2026. The ISO link is a public overview, not a claim of access to its full paid text. EURAMET cg-4 and cg-14 provide force and static-torque context; neither supplies a universal six-axis qualification rule. Scores, thresholds, tables and scenarios are editorial screening aids. No laboratory scope, commercial price or HexaForce product capability is verified by this page. Confirm the current schedule, method, range and actual calibration uncertainty before contracting.

Calibration suitability depends on the complete measurement method, traceability, uncertainty and application requirements. To audit an incoming calibration certificate or verify uncertainty budgets, visit 6-axis force sensor calibration certificate review or determine recalibration schedules with the calibration interval decision worksheet.